SOP-Antibiotic Zone Reader

SOP-Antibiotic Zone Reader

ANTIBIOTIC ZONE READER
STANDARD OPERATING PROCEDURE

  1. Connect the three pin plug to the main supply.
  2. Switch “ON” the main switch. Digital display will be showing some reading.
  3. Switch “ON” the Lamp.
  4. Turn the right side ‘Slider knob’ fully anticlockwise.
  5. Keep the sample petriplate on the top belt.
  6. Put magnetic clamp around the petriplate and slide it manually.
  7. One edge of circle zone image just touches the mark line on the viewing glass.
  8. Rotate the dial given below the display unit to focus the image of the sample.
  9. If required, again adjust the arc precisely from one edge of zone, tangential with mark line on the viewing glass by the ‘slider knob’.
  10. Press “ZERO” switch for few seconds and then release to make the reading 00.0 on the display.
  11. Now turn the right side slider knob slowly, in clock wise position, till the other edge of the zone image touches the mark line on the viewing glass.
  12. Note the reading in millimeter (mm) directly on the digital display.

Calibration:

  1. Take the calibrated circular disc.
  2. Place the disc between the magnetic holders.
  3. Adjust scale to zero and move the scale so that the entire disc is covered.
  4. Record the data in calibration record format.
  5. Measure the diameter of the disc by Vernier caliper and record the reading in calibration record format.
  6. Acceptance criteria: ±2 mm of measured diameter.

Leave a Reply

Your email address will not be published. Required fields are marked *

×
Free Video Lectures of Pharmacy Exams
Watch now
Admission open 2024 Clinical Research & Pharmacovigilance Courses
Apply now

Developed By Connect Globes