VALIDATION OF UV SPECTROPHOTOMETER INSTRUMENT: PROCEDURE & TEST
VALIDATION OF UV INSTRUMENT :-
Performance test
1. Wavelength accuracy :-
Standard :- 4% holmium oxide in 10% perchloric acid.
Specification :- NMT _+ 1 nm in 200- 400nm. NMT _+ 3 nm in 400- 600nm.
2. Stray light :-
Standard :- 1.2% w/v potassium chloride solution.
Specification :- Absorbance at 200nm should be more than 2.
3. Resolution
Standard :- 0.02 % v/v toluene in hexane solution.
Specification :- The ratio of the absorbance at λmax (269nm) and Absorbance at λmax(266nm) should be greater than 1.5
4. Photo metrics accuracy :-
Standard :- potassium dichromate solution in KOH.
5. Cell thickness tolerance :-
Standard :- for 1 cm
Specification :- Tolerance allowed is 0.005 cm.